OdaLog® Low Range H2S Logger
The OdaLog Low Range H2S Logger is a high sensitivity, convenient,
weather resistant, and portable instrument designed to assist in H2S
odor management activities by accurately measuring concentrations
of H2S much lower than the standard range of OdaLog H2S Loggers. With
a range of 0.01 to 2.00ppm, the OdaLog Low Range H2S Logger allows
reliable measurement of H2S based odours at plant fence-lines, scrubber
outlets, land fill sites, and residential locations where odour complaints
have been reported.
Typically the OdaLog Low Range H2S Logger may be placed at the outlet
of a scrubber for performance tracking or several units may be mounted
on a fence line or plant perimeter for the purpose of measuring concentrations
of H2S leaving a facility and entering adjoining properties.
Data can be logged at sampling intervals of between 10 minutes and
one hour. The recorded H2S data can be downloaded to a computer using
the standard OdaLog infrared link and OdaLog software. Detailed information
on the OdaLog software can be found on the OdaLog Low Range H2S Logger
website at http://www.odalog.com/OL_Software.htm
An internal battery (1 x standard C size Alkaline cell) will operate
the built-in sampling system continuously for at least seven days
and a separate internal Lithium cell powers the main processor functions
for at least two months.
Specifications (Preliminary) - OdaLog® Low Range H2S Logger
| Measurement
Range: |
0.01 to 2.00ppm H2S |
| Zero Drift (NTP): |
+/- 0.01ppm Conditions: NTP, fresh air,
taken over 10 consecutive sample cycles |
| Precision: |
5% Relative Standard Deviation Conditions:
NTP, 0.20ppm H2S applied, taken over 10 consecutive
sample cycles. |
| Accuracy: |
+/- 10% of reading 0.10 to 2.00ppm Conditions:
NTP, 0.20ppm applied, taken over 10 consecutive
sample cycles |
| Environmental
Protection: |
IP54 (dust and splash protected) Uses OdaLog
double O-ring sealing technology |
| Instrument Temperature
Range: |
0° C to 40° C and 6°C change / Hr
maximum
(32° F to 104° F and 11° F change / Hr maximum) |
| Logging/sampling
interval: |
10 minutes to 1 hour |
| Memory capacity: |
30000 data points |
| Relative Humidity
Range: |
15-90% (non-condensing) |
| Pressure Range: |
Atmospheric ±10% |
| External Dimensions: |
62mm (2.44") diameter X 307mm (12.1")
long |
| Weight: |
Approximately 900grams (2.0lb) |
| Battery life
(and type): |
CPU battery: 2 Months (2/3AA size Lithium cell)
Pump battery: 7 Days (C size Alkaline cell) |
| Warranty: |
12 Months |